Specifications

Flow Rare & Accuracy
701-1500ml/hr1+x)% of reading where x = 0.0033% x (flow rate -700ml/hr)
10-700ml/hr1% of reading
0.1-9.9ml/hr1% of reading + 0.005 ml/hr
Volume0.001 to 9999ml
Flow Update Rate1Hz
PCA / Bolus Measurements (Volume)
Display Range0.1 to 100ml
Measuring Range0.5 to 100m
Accuracy± 1% of the reading
Max. Resolution10μl
Basal Flow Rate1ml to 30ml/h
PressureMax. 2500mmHg
Flow Measurement
Display Range0.010 to 1500ml/h
Max. Display Resolution>10μl/h
Measured Range0.100 to 1450ml/h
Occlusion/Back Pressure Measurement
Pressure Measurement Range-500 to 2500mmHg
Back Pressure Setting Range-200 to 600mmHg
Accuracy-500 to 1000mmHg ± 10mmHg[br] 1000 to 2500mmHg ± 1% of reading
UnitmmHg, PSI, Bar, kPa
Resolution1mmHg
General Specifications
Storage15 x 24h of testing at 1 second sampling rate
Dimensions300 x 204 x 220mm/11.8 x 8 x 8.6” (w x d x h)
Weight5kg/11lbs (1 channel)[br]6kg/13lbs (2 channel)[br]8kg/18lbs (4 channel)
Mains Supply90-264VAC, 50/60 Hz, 60W
Mains Cable1ml to 30ml/h
Storage Environment0–50°C / 32–122°F
Operating Conditions15–40°C / 59–104°F
Environmental ProtectionIP40
PC CommunicationUSB B
Keyboard CommunicationUSB A
DisplayLCD color graphic display ¼” VGA