X
Flow Rare & Accuracy | |
---|---|
701-1500ml/hr | 1+x)% of reading where x = 0.0033% x (flow rate -700ml/hr) |
10-700ml/hr | 1% of reading |
0.1-9.9ml/hr | 1% of reading + 0.005 ml/hr |
Volume | 0.001 to 9999ml |
Flow Update Rate | 1Hz |
PCA / Bolus Measurements (Volume) | |
---|---|
Display Range | 0.1 to 100ml |
Measuring Range | 0.5 to 100m |
Accuracy | ± 1% of the reading |
Max. Resolution | 10μl |
Basal Flow Rate | 1ml to 30ml/h |
Pressure | Max. 2500mmHg |
Flow Measurement | |
---|---|
Display Range | 0.010 to 1500ml/h |
Max. Display Resolution | >10μl/h |
Measured Range | 0.100 to 1450ml/h |
Occlusion/Back Pressure Measurement | |
---|---|
Pressure Measurement Range | -500 to 2500mmHg |
Back Pressure Setting Range | -200 to 600mmHg |
Accuracy | -500 to 1000mmHg ± 10mmHg[br] 1000 to 2500mmHg ± 1% of reading |
Unit | mmHg, PSI, Bar, kPa |
Resolution | 1mmHg |
General Specifications | |
---|---|
Storage | 15 x 24h of testing at 1 second sampling rate |
Dimensions | 300 x 204 x 220mm/11.8 x 8 x 8.6” (w x d x h) |
Weight | 5kg/11lbs (1 channel)[br]6kg/13lbs (2 channel)[br]8kg/18lbs (4 channel) |
Mains Supply | 90-264VAC, 50/60 Hz, 60W |
Mains Cable | 1ml to 30ml/h |
Storage Environment | 0–50°C / 32–122°F |
Operating Conditions | 15–40°C / 59–104°F |
Environmental Protection | IP40 |
PC Communication | USB B |
Keyboard Communication | USB A |
Display | LCD color graphic display ¼” VGA |