Oxygen Saturation | |
---|---|
Range | 50 to 100% |
Accuracy | ± 1 bpm |
Chronometer | Test option to test response time SpO2 monitor |
Arrhythmia Waveforms | |
---|---|
Atrial | Sinus Arrhythmia (SA), Missing beat, Atrial Flutter (AFLT), Atrial Fibrillation (AFB), Paroxysmal Atrial Tachycardia (PAT), Junctional Premature Contraction |
Atrial Conduction | First Degree AV Block, Second Degree AV Block - Mobitz I, Second Degree AV Block - Mobitz II, Third Degree AV Block, Right Bundle Branch Block (RBB), Left Bundle Branch Block (LBB), Left Anterior Hemiblock |
Ventricular | Premature Ventricular Contraction - Intermittent Premature Ventricular Contraction – Continuous, Bigeminy, Trigeminy, Ventricular Flutter (VFLT), Ventricular Fibrillation (VFB), Ventricular Tachycardia (VTC), Right Focal PVC |
Respiration | |
---|---|
Rates | 5, 10, 15, 30, 60, 120, 180 Breaths per Minute |
Base Resistances | 250, 500, 750, 1000 ohms |
Apnoea Simulation | 0 – 60 seconds duration 0 – 300 seconds interval |
General Specification | |
---|---|
Operation | Battery cell, insitu charge |
Battery Charger | 110-230VAC, 50/60 Hz |
Battery Life | 4-24 hours testing depending on simulations |
Memory Capacity | Approximately 10,000 records |
Communication | Via Blue Tooth |
Display | Monochorme, ¼ VGA full graphics |
Keypad | Alpha-numeric |
Weight | <1.5 kg |
Size (L x W x D) | 270 x 110 x 75 mm / 10.5 x 4 x 3” |
Operating Conditions | 10-30°C, 0-90% RH - NC |
Storage Environment | -15°C - +60°CC |
Environmental Protection | IP 40 |
Non-invasive Blood Pressure | |
---|---|
Simulation Waveform | Oscillometric |
Integrated Pump | 0 to 350 mmHg user configurable |
Leak Test | User configurable between 0-350mmHg |
Chronometer | Configurable upto 999 secs |
Digital Manometer | 0 - 410 mmHg |
Pressure Accuracy | +/- 0.5% FS |
Pressure Units | mmHg, inHg, kg/cm2, cmH2O, mBar, PSI, inH2O and kPa |
ECG Arrhythmia Simulator | |
---|---|
ECG | 5 lead simulation including high level output |
Wave Forms | Normal Sinus Rhythm (NSR), ST Elevation, ST Depression, Myocardial Infarction, Tall T |
Rate | 20 – 300 BPM typed value |
Performance Waveforms | Sine, Square, Triangle, and Pulse |
Pacer Waveforms | Synchronous Atrial, Asynchronous Atrial |
Pacer Only | Pacer Pulse rate 60 BPM, Ventricular Pacer rate 70 BPM, Atrial & Ventricular Pacer Heart rate 70 BPM, R Wave Detection |
Temperature | |
---|---|
Simulation | YSI 400 / 700 Static |
Range | Preset at 25, 33, 37and 41°C |
Invasive Blood Pressure | |
---|---|
Channels | 2 channels |
Static | 0 to 300mmHg. Typed values |
Dynamic | 0 - 300mmHg for systolic & diastolic |