Safetest 60+

SPECIFICATIONS & MANUALS

If you're looking for the Safetest 60+ datasheet featuring full technical specifications or the instruction manual, you can find them here.

Specifications

Insulation Resistance
MeasurementPoint-to-point and EUT to Earth
Voltage500VDC @1mA
Maximum O/C Voltage<600VDC
Range100K – 100MΩ ± 5% ± 2 Digits
Resolution0.01MΩ
Short Circuit Current<2mA
Voltage Measurement
ApplicationL-N, L-E, N-E and touch voltage (IEC 61010)
Range0.0V – 300VAC
Resolution0.1Hz
AccuracyUnspecified
Mains frequency45.0 – 66.0Hz
Power Source
Maximum Current Rating20A @ 120V / 16A @ 230V
Duty cycle (@21°C ambient)16A to 20A, 3 min. on/ 10 min. off 10A to 15A, 3 min. on/ 5 min. off 0A to 10A, continuous
Mains Power90-264V 48-64Hz
Earth Continuity
Pre-pulse2 wire ZAP technique
Pulse ShapeExponential Decay
Decay Time200 – 550μs to 5% of peak (0.1 to 0.8Ω respectively)
MethodWire
Measurement Current>± 200mADC into 2Ω
Max Test Voltage4-24Vrms o/c
Measuring Range (Low Range)0.001 – 0.999Ω
Resolution0.1Ω
Measuring Range (Mid Range)1.00 – 9.99Ω
Measuring Range (High Range)10.0 – 19.9Ω
Accuracy± 1% of value, ± 5mΩ ± 1% of value, ± 5mΩ
Circuit ProtectionTest inhibited if ≥ 30VAC or DC at 4mm inputs
Powered Leakage Measurements
IEC 62353Equipment Leakage (Direct)
IEC / AAMI 60601Earth + Enclosure Leakage + Patient Leakage
NFPA-99Ground + Chassis Leakage + Patient Leakage
IEC 61010Touch Leakage, Touch Voltage
Test VoltageMains Supply Voltage
Measuring Range0.1 - 9999μA (0.1 - 8000μA typical for IEC61010)
Measurement/Display Resolution0.1μA
Accuracy± 2%, ± 5μA
Mains ReversalSoft Key
Single Fault ConditionsOpen Neutral and Open Earth Via Soft Key
Frequency ResponseIEC 60601 – 62353, NFPA-99, and IEC 61010 selectable
EUT Load Current Measurement
Range0.0A – 20.0A
Resolution0.1A
Accuracy±5% ± 2 digits
General Specifications
Weight1.1kg/2lbs (unit) 2.2kg/5lbs (complete with accessories)
Size (L x W x D)225 x 150 x 100mm / 9 x 6 x 4”
Operating Temperature0 to 40°C
Humidity0 – 98% Relative humidity, non-condensating
Storage Temperature-10 to 50°C
Altitude0 – 2000m
Ingress ProtectionIP 40
Operating Pollution Degree2, according to IEC 60529
Overvoltage CategoryCAT II 300V